ASTM F1089

Corrosion of surgical instruments

ASTM F1801

Corrosion fatigue testing of metallic implant materials

Icon of the ASTM F1875 test

ASTM F1875

Fretting corrosion testing of modular implants – Femoral stem-head interface

ASTM F2129

Corrosion test of small implants

ASTM F3044

Galvanic corrosion

ASTM F897

Fretting corrosion of osteosynthesis plates and screws

ASTM G102

Calculation of corrosion rates

ISO 13402

Surgical and dental hand instruments – Determination of resistance against autoclaving, corrosion and thermal exposure

ISO 16429

Assessment of corrosion behaviour of metallic implantable materials

ASTM F1875

Fretting corrosion testing of modular implants – Femoral stem-head interface

Normative References

ASTM F1875: Standard Practice for Fretting Corrosion Testing of Modular Implant Interfaces: Hip Femoral Head-Bore and Cone Taper Interface

Two methods are part of this standard, a quantitative analysis (method I; long-term evaluation) and a qualitative assessment (method IIb; short-term evaluation).

Method I is used to determine the fretting corrosion behavior in the long-term. The interface surface damage and generation of corrosion products are determined through optical evaluation (micro- and macroscopic, SEM/EDX) and particle and ion concentration analyses of the test solution after cyclic loading.

Method IIb is a short-term electrochemical evaluation (current measurement) to compare different designs and surface treatments. This requires extremely precise testing equipment and a well-shielded test environment.

This standard defines the testing, analytical, and characterization methods for evaluating the fretting corrosion performance at the bore and cone interface of the head-stem junction in modular hip implants under cyclic loading in a corrosive environment.

What is the difference between Method I and Method IIb under ASTM F1875?

Method I = long-term, cyclic loading with fluid analysis and surface characterization to quantify debris/ions and damage.
Method IIb = short-term electrochemical (current measurement) screening under cyclic loading for comparative assessment (for design changes only).

Which method under ASTM F1875 is needed for regulatory approval?

Typically, the long-term method I is expected for submissions. Method IIb is intended to compare design changes. Always confirm with your regulatory authority and submission strategy.

When do I need ASTM F1875?

Whenever your THR design includes a modular femoral head–neck junction, and you need evidence on fretting/fretting-corrosion at the taper interface. It’s also referenced within ISO 21535’s normative set.

What are typical observations for devices tested under Method I of ASTM F1875?

Corrosion, discoloration or debris on the taper surfaces and dulling of machine lines as well as particle and ion release into the test fluid.

What other tests are typically conducted in conjunction with ASTM F1875?

Post-Fatigue: Static Disassembly Test according to ISO 7206-10 or ASTM F2009 and particle analysis according to ASTM F1877 and ASTM F561

Is there an acceptance criterion for ASTM F1875?

There is no acceptance criterion defined in the standard. Predicate device testing under the same conditions is recommended.

The long-term method is required for approval.

Resources

Contact person Stephan Herdegen

Your contact person

Stephan Herdegen, M.Sc.

Technical Specialist Corrosion