ISO 25539

ISO 10555-3

Central venous catheters

ISO 10555-4

Balloon dilation catheters

ISO 10555-5

Over-needle peripheral catheters

ISO 80369-20

Small-bore connectors test methods

Simulated use

Normative References

ISO 25539-1: Cardiovascular implants – Endovascular devices – Part 1: Endovascular prostheses

ISO 25539-2: Cardiovascular implants – Endovascular devices – Part 2: Vascular stents.

These tests determine the delivery system’s ability to be pushed or positioned by an operator without bending or buckling. We evaluate the ease of advancing the delivery system into the model (trackability), the ability of the delivery system to be removed from the model (withdrawability) and the ability to transmit torque from the proximal end to the distal end of the catheter (torquability), if applicable. To evaluate trackability and withdrawability, a universal test frame is used. Torquability is tested in a custom test set-up. The vascular model in use is described by ASTM F2394.  