ASTM F2068: Standard Specification for Femoral Prostheses – Metallic Implants
ISO 7206-4: Implants for surgery – Partial and total hip joint prostheses – Part 4: Determination of endurance properties and performance of stemmed femoral components
ISO 7206-6: Implants for surgery – Partial and total hip joint prostheses – Part 6: Endurance properties testing and performance requirements of neck region of stemmed femoral components
ASTM F1044: Standard Test Method for Shear Testing of Calcium Phosphate Coatings and Metallic Coatings
ASTM F1147: Standard Test Method for Tension Testing of Calcium Phosphate and Metallic Coatings
ASTM F1978: Standard Test Method for Measuring Abrasion Resistance of Metallic Thermal Spray Coatings by Using the Taber Abraser
The standard describes four different types of femoral stems and summarizes several considerations concerning the fatigue performance of femoral hip components and the integrity of their metallic coatings.
Fatigue performance can be assessed using the following standards:
- ISO 7206-4: Femoral stem fatigue test
- ISO 7206-6: Femoral neck fatigue test
Coating integrity can be investigated regarding shear strength, tensile strength and – for plasma sprayed thermal coatings – abrasion resistance. Standards and minimum acceptance criteria are as below:
- ASTM F1044 – Shear strength – average equal to or exceeding 20 MPa
- ASTM F1147 – Tensile strength – average equal to or exceeding 20 MPa
- ASTM F1978 – Abrasion resistance – average ≤ 65 mg/100 cycles
ASTM F2068 was withdrawn with no replacement. We have transferred the ASTM standard into an internal, accredited testing standard (PI-98).
EndoLab is one of the most established implant testing laboratories worldwide and performs testing according to ASTM F2068 (internal method PI-98) in its ISO 17025 accredited laboratory. As active contributors to international standardization, our engineers know not only the standard, but also the reasoning behind it.
EXECUTIVE SUMMARY
ASTM F2068 sets requirements for metallic stemmed femoral prostheses for hemi-arthroplasty and total hip arthroplasty. Fatigue samples should represent the worst-case stress condition of the design series, tested with the worst-case offset head. The stem is loaded cyclically according to ISO 7206-4, the neck region according to ISO 7206-6. Porous metal coatings are tested for shear strength, tensile strength and abrasion resistance. The implant must withstand the cyclic loading and the coatings must reach the minimum values. Alternatively, fatigue strength may be shown to equal that of a comparable, clinically successful design.
Read more…
Key facts – ASTM F2068
- Last version: ASTM F2068-15 (withdrawn 2023); continued at EndoLab as accredited method PI-98
- Stem fatigue: ISO 7206-4:2010, with worst-case offset head
- Neck fatigue: ISO 7206-6:2013 with torsion, 534 N to 5340 N, 10 million cycles
- Coating shear strength (ASTM F1044): average ≥ 20 MPa
- Coating tensile strength (ASTM F1147): average ≥ 20 MPa
- Abrasion of plasma spray coatings (ASTM F1978): average ≤ 65 mg/100 cycles
- Porous surface (typical): 30–70 % porosity, 100–1000 µm pore size, 500–1500 µm thick
Purpose and scope of ASTM F2068
ASTM F2068 covers metallic femoral stems with an integral head or with a cone for a modular head, fixed by press fit, bone cement or bone ingrowth. ASTM F2068 does not cover:
- modular femoral heads themselves
- requirements for modular connections of multi-component stems
- custom implants designed for a single patient
- non-metallic implants, non-porous bioactive ceramic or porous polymer coatings
Its requirements are based on more than forty years of clinical experience.
Materials in ASTM F2068
ASTM F2068 lists titanium, cobalt-based and stainless steel materials that have been used successfully. Material conformance and earlier clinical use do not ensure implant strength, as design and manufacturing strongly influence material properties. Every implant must meet the performance requirements. Materials with little clinical history must show corrosion resistance (ASTM F746) and biological response (ASTM F748, ASTM F981) at least equal to a listed material.
Fatigue performance according to ASTM F2068
Stems must withstand normal physiological static and dynamic loads without overload fracture, plastic deformation or fatigue fracture. According to ASTM F2068, the worst case is not automatically the longest offset head: stress depends on offset combined with the cross-section near the cantilever plane. ASTM F2996 (finite element analysis) may help identify it. ISO 7206-4 and ISO 7206-6 test modular stems in saline and non-modular stems in air. Environmentally sensitive materials should be tested in a simulated physiological environment.
Coating integrity in ASTM F2068
In ASTM F2068, porous metal coatings such as plasma spray or fibre metal must allow tissue ingrowth for firm fixation. The shear and tensile limits derive from FDA guidance and are accepted as reasonable lower bounds.
This page is EndoLab’s summary of ASTM F2068 and does not replace the standard itself. The complete and authoritative version, ASTM F2068-15 (withdrawn 2023), is available from ASTM International: ASTM F2068-15