ASTM F2193

ASTM F1717

Spine corpectomy model

ASTM F1798

Spinal implants - interconnections mechanisms and subassemblies

ASTM F2077

Spinal cage test

ASTM F2193

Spinal components

ASTM F2267

Spinal implants subsidence test

ASTM F2346

Spinal disc static and fatigue test

ASTM F2624

Extra discal motion preserving implants

ASTM F2694 / F2790

Lumbar total facet prostheses

ASTM F2706

Occipital-cervical and occipital-cervical-thoracic model

ASTM F2789

Mechanical and functional characterization of nucleus devices

ASTM F3295

Impingement wear testing of total disc prostheses

ASTM F3631

Cage impact test

ISO 12189

Spine corpectomy model with anterior support

ISO 18192-1

Spinal disc wear test

ISO 18192-2

Nucleus replacement wear test

ISO 18192-3

Impingement-wear testing of lumbar prostheses

ISO 21534

Particular requirements for joint replacement implants

ISO 23089-2

Spinal cage pre-clinical mechanical assessment

PI-52

Expulsion test

PI-76

Spinous process plates

Spinal components

Normative References

ASTM F2193: Standard Specifications and Test Methods for Components Used in the Surgical Fixation of the Spinal Skeletal System.

Specifications and test methods for metallic spinal screws, metallic spinal plates and metallic spinal rods is given within this standard. If the user is interested in evaluating the next level in the spinal fixation system chain, the interconnections between individual components and subassemblies (two or more components), the user should consult ASTM F1798. At the highest level within this chain is the test method ASTM F1717, which is used to evaluate an entire construct assembled from many components and involves numerous interconnections and several subassemblies.  

Contact Person male

Your contact person

Dipl. Ing. Christian Findeiss

Head of Fatigue Testing & Material Analysis, Head of Vascular Testing