ASTM F1820

ASTM F1820

Disassembly forces of modular acetabular devices

ASTM F1875

Fretting corrosion testing of modular implants – Femoral stem-head interface

ASTM F2009

Modular connection disassembly test

ASTM F2068 (PI-98)

Requirements for femoral prostheses - Metallic implants

ASTM F2345

Femoral head fatigue test

ASTM F2580

Modular stem fatigue test

ASTM F2582

Impingement test

ASTM F2979

Bearing surface characterization of retrieved hard-on-hard hip prostheses

ASTM F2996

FEA non-modular hip femoral stem

ASTM F3018

Guide for assessment of hard-on-hard articulation THR devices

ASTM F3047M

High demand hip wear testing

ASTM F3090

Fatigue test of acetabular devices

ASTM F3446

3D frictional torque measurement of total hip joint prostheses

ISO 11491

Impact resistance femoral heads

ISO 14242-1

Hip wear test

ISO 14242-1

Third body wear test based on ISO 14242-1

ISO 14242-4

Microseparation test

ISO 21534

Particular requirements for joint replacement implants

ISO 21535

Hip range of motion

ISO 21535

Specific requirements for hip-joint replacement implants

ISO 7206-1

Classification and designation of dimensions

ISO 7206-10

Femoral head compression test

ISO 7206-12

Deformation test for acetabular shells

ISO 7206-13

Femoral head torsion test

ISO 7206-2

Hip surface analysis

ISO 7206-4

Femoral stem fatigue test

ISO 7206-6

Femoral stem neck fatigue test

ISO 7206-8

Femoral stem fatigue test

PI-11

Insert compression fatigue / static test

PI-3

Acetabular cup luxation test

PI-58 (ISO 7206-8)

Femoral stem fatigue test

PI-87

FEA non-modular hip femoral stem - neck region

PI-99 (ASTM F2091)

Standard specification for acetabular prostheses

Disassembly forces of modular acetabular devices

Normative References

ASTM F1820: Standard Test Method for Determining the Forces for Disassembly of Modular Acetabular Devices

This test method covers three different tests to determine the mechanical strength of the fitting between the acetabular shell and liner.

The push-out test determines the axial locking strength of the acetabular liner within the acetabular shell.
The offset pull-out or lever-out disassembly test determines the resistance of the locking mechanism to edge forces. They could occur when the neck of the femoral stem impinges on the edge of the acetabular liner.
The torque-out test determines the resistance of the locking mechanism against torsional forces.  