ISO 25539

ISO 10555-3

Central venous catheters

ISO 10555-4

Balloon dilation catheters

ISO 10555-5

Over-needle peripheral catheters

ISO 80369-20

Small-bore connectors test methods

Force to deploy

Normative References

ISO 25539‐1: Cardiovascular implants – Endovascular devices – Part 1: Endovascular prostheses

ISO 25539‐2: Cardiovascular implants – Endovascular devices – Part 2: Vascular stents

EndoLab® determines the force to deploy a self-expanding stent. All applicable steps of the deployment process will be evaluated. The stent system is inserted in an anatomical model and the deployment is initiated and completed at a rate that simulates clinical use while measuring the force to deploy the self-expanding stent.  

Contact Person male

Your contact person

Dipl. Ing. Christian Findeiss

Head of Fatigue Testing & Material Analysis, Head of Vascular Testing