PI-53

ASTM F1223

Knee constraint test

ASTM F1672 (PI-95)

Patella specification

ASTM F1814

Test of modular knee implants

ASTM F2009

Modular connection disassembly test

ASTM F2083 (PI-96)

Knee specification

ASTM F2722

Fatigue test baseplate/bearing fixation under torsional loading

ASTM F2723

Fatigue test baseplate/bearing fixation

ASTM F2724

Dislocation mobile bearing

ASTM F2777

Fatigue test under high flexion

ASTM F3140

Unicondylar tibial tray fatigue test

ASTM F3141

Knee loading profiles

ASTM F3161

FEA knee femoral component

ASTM F3210 / PI-73

Knee condyle fatigue test

ASTM F3334

FEA Tibial component

ASTM F3495

Ceramic knee failure load

ISO 14243-1/3

Knee wear test

ISO 14243-5

Durability performance patellofemoral joint

ISO 14879-1 / ASTM F1800

Tibial tray fatigue test

ISO 21534

Particular requirements for joint replacement implants

ISO 21536

Specific requirements for knee-joint replacement implants

ISO 7207-2

Knee surface analysis

PI-17

Knee contact pressure test

PI-53

Stemmed femoral knee component fatigue test

Stemmed femoral knee component fatigue test

Normative References

PI-53: Fatigue testing of stemmed femoral TKA components

The purpose of the test described in this internal test procedure is to determine the mechanical strength of stemmed femoral TKA components loaded under assumed worst case conditions.
The femoral component and the stem have to be assembled according to the customer´s instructions and tightened with a calibrated torque wrench (if required).

The load has to be applied at an angle of 25° to the neutral position of the femoral component as indicated by the client. This simulates a slightly flexed knee position. A varus malalignment has to be simulated by a 5° inclination of the specimen holder. This alignment leads to load only the medial condyle.
The suggested embedding level of the stem (if present) is 80 mm.

A load of 2,000 N (R=1/10) at a rate of ≤ 10 Hz is suggested to be used for the dynamic tests. Modular implants have to be tested in Ringer® solution at 37 °C.
The run-out criteria has to be set to 10,000,000 cycles.  

Contact Person male

Your contact person

Dipl. Ing. Christian Findeiss

Head of Fatigue Testing & Material Analysis, Head of Vascular Testing