ASTM F2129

ISO 10555-3

Central venous catheters

ISO 10555-4

Balloon dilation catheters

ISO 10555-5

Over-needle peripheral catheters

ISO 80369-20

Small-bore connectors test methods

Corrosion test of small implants

Normative References

ASTM F2129: Standard Test Method for Conducting Cyclic Potentiodynamic Polarization Measurements to Determine the Corrosion Susceptibility of Small Implant Devices

This test method evaluates the corrosion susceptibility of small medical implant devices, single device components or other material samples made of metal. The test is performed using potentiodynamic polarization in a cyclic manner (forward and reverse). For measurement, a high-precision computerized potentiostat is used.
Small medical implant devices covered by this standard include but are not limited to, stents (vascular and ureteral), filters, grafts, occluders, staples and aneurysm or ligation clips. Phosphate buffered saline (PBS) is used as the standard test fluid, but other simulated physiological solutions are also possible.

Your contact person

Stephan Herdegen, M.Sc.

Technical Specialist Corrosion