ISO 14243-5

ASTM F1223

Knee constraint test

ASTM F1672 (PI-95)

Patella specification

ASTM F1814

Test of modular knee implants

ASTM F2009

Modular connection disassembly test

ASTM F2722

Fatigue test baseplate/bearing fixation under torsional loading

ASTM F2723

Fatigue test baseplate/bearing fixation

ASTM F2724

Dislocation mobile bearing

ASTM F2777

Fatigue test under high flexion

ASTM F3140

Unicondylar tibial tray fatigue test

ASTM F3141

Knee loading profiles

ASTM F3161

FEA knee femoral component

ASTM F3210 / PI-73

Knee condyle fatigue test

ASTM F3334

FEA Tibial component

ASTM F3495

Ceramic knee failure load

ISO 14243-1/3

Knee wear test

ISO 14243-5

Durability performance patellofemoral joint

ISO 14879-1 / ASTM F1800

Tibial tray fatigue test

ISO 21534

Particular requirements for joint replacement implants

ISO 21536

Specific requirements for knee-joint replacement implants

ISO 7207-2

Knee surface analysis

PI-17

Knee contact pressure test

PI-53

Stemmed femoral knee component fatigue test

PI-96 (ASTM F2083)

Knee specification

Durability performance patellofemoral joint

Normative References

ISO 14243-5: Implants for surgery – Wear of total knee prostheses – Part5: Durability performance of the patellofemoral joint – Loading and displacement parameters for testing machines and corresponding environmental conditions for test

Loading and displacement waveforms, which simulate the patellofemoral kinetics and kinematics by combining a squatting sub-cycle with two walking gait sub-cycles, are performed until delamination, cracking, or another identified failure mode occurs.

If the implants do not demonstrate any failure modes by 50 thousand cycles, the test is stopped. At least five polymeric knee joint patellar components are tested. Prior to testing the patellar components are artificially aged according to ASTM F2003.

The EndoLab® knee joint simulator is used for testing. Please download our video to see the simulator moving.  

Your contact person

Jason Steffens, M.Sc.

Head of Tribology, Head of FEA