EndoLab - Implant Testing Services Implant Testing

Small bone plates - PI-68

Normative References  



PI-68: (WK52025) Standard Test Method for Metallic Bone Plates Used in Small Bone Fracture Fixation

This standard is used for testing of small bone metallic plates. Small bone plates are used primarily in the far extremities (e.g. fingers and toes) and in the craniomaxillofacial region of the body for fracture fixation. Test standards for metallic bone plates, ASTM F382 and ISO 9585, currently exist, but due to the smaller size of the plates used in small bone fracture fixation, some of the tests described in these standards are very difficult to set up and execute. This standard describes static and dynamic tests by specifying load types and specific methods of applying these loads. The following tests to evaluate and characterize them can be carried out at EndoLab®: Static and dynamic torsion, static and dynamic cantilever beam bending, static and dynamic lateral bending, and static and dynamic three-point bending.  


Implant Testing - Small bone plates - PI-68